Aeroneb Go Troubleshooting

Aerogen
  1. There is no visible aerosol from the nebulizer when using batteries

  2. No visible aerosol when using AC/DC Adapter

  3. No visible aerosol when power source is functioning

  1. There is no visible aerosol from the nebulizer when using batteries

Check that the batteries are inserted correctly. Using the markings under each battery location, place each battery in the Control Module housing. Match (+) and (-).

Check to see if the low battery power light is flashing. Replace the batteries or use the AC/DC Adapter.

Check that the Control Module cable is connected properly to the nebulizer and to the Control Module. Connect the cable properly to the nebulizer and the Control Module

  1. No visible aerosol when using AC/DC Adapter

Check that the AC/DC Adapter is plugged correctly into a working AC outlet and the green light is on the AC/DC Adapter. Insert the plug into a working outlet and verify that the green light on the controller is lit

Check that the AC/DC Adapter cable is connected properly to the Control Module. Connect the AC/DC Adapter cable properly to the Control Module.

Check that the AC/DC Adapter is not faulty. Try to operate the Control Module by using batteries instead of the AC/DC Adapter. If the Control Module functions properly with batteries but not with the AC/DC adapter, replace the AC/DC Adapter.

  1. No visible aerosol when power source is functioning

Check that there is medication in the Medication Cup. Fill the Medication Cup with prescribed medication.

Check that the nebulizer has been properly cleaned.